High low temperature/ low air pressure test chamber
This chamber is designed to test products and materials to the effects of high and low temperature (and humidity), and low atmospheric pressure.
The choice of equipment depends on many factors:
◆ The size and weight of the test product;
◆ Requirements of test modes (graphics) - complexity, speed of changes in temperature and / or humidity parameters, pumping time, duration-regulated by Russian or other standards, including international standards;
◆ The presence of heat from the test product;
◆ Individual technical requirements, for example, specifications of a specific product.
Technical specifications:
Model | QD701HD | QD702HD | QD705HD | QD710HD |
Internal volume | 120 L | 180 L | 504 L | 1000 L |
Internal dimensions (D*W*H) mm | 400*500*600 mm | 500*600*600 mm | 700*800*900 mm | 1000*1000*1000 mm |
External dimensions (D*W*H) mm | 1900*800* 1550 mm | 2050*900* 1550 mm | 3250*1150* 1950 mm | 3400*1450*1950 mm |
Temperature Control mode | BTPC: Balanced temperature & pressure control |
Working environment temperature | +5…+35 ℃ |
Performance |
Temperature range* | -70℃...+150℃ (high temperature: 180℃, optional); |
Pressure range | Atmospheric pressure... 0.5 kPa |
Temperature fluctuation | ≤±0.5℃ (under atmospheric pressure, no-load) |
Temperature uniformity | ≤2℃ (under atmospheric pressure, no-load) |
Temperature deviation | ≤±2℃ (under atmospheric pressure, no-load) |
Pressure accuracy | Atmospheric pressure~40KPa: ±2 KPa; 40KPa~10KPa :±5%; |
Heating time | ≤60 min. (25℃…+100℃) |
Cooling time | ≤90 min. (+25℃…-70℃) |
Pressure recovery rate | ≤1 kPa/min |
Depressurization time | Atmospheric pressure~10KPa, ≤30min; |
Material |
Shell material | Cold rolled sheet steel, spray plastics on surface |
Internal material | Stainless steel |
Thermal insulation material | Polyurethane foam + glass wool |
Refrigeration system |
Refrigeration mode | Mechanical compression, two-stage refrigeration |
Compressor | Imported enclosed/ semi-enclosed compressor |
Cooling mode | Air cooling/ water cooling |
Heater | Stainless steel heater |
Blower | Centrifugal blower |
Vacuum pump | Leybold, Germany |
Temperature sensor | PT100 |
Controller | Touch screen, programmable control; |
Power supply | AC380V, 50Hz, three-phase four-wire + earth wire; |
*Note:
temperature range -40℃~+150℃: QD4**HD series;
High-low temperature low air-pressure test chamber with humidity from 20 to 98% R.H .: THP *** HD series.