Testing chamber

UV-resistant Climate test chamber

2012-6-6 14:34:19



Model

HD-UV-1/HD-UV-2

Internal dimensions

45x110x50cm

Standard

GB/T16422.3,GB/T5170.9

Performance

Temperature range

R.T. +10~70

Temperature fluctuation

≤±0.5

Humidity range

95%~98%RH

External dimensions D× W× H

50x130x148cm

Working environment temperature

+5~+35

Light source

UV-A/B light

 Light source wavelength

280~400nm

Size of sample holder

150x75x1.5mm

Center distance between sample and lamp

50mm±2mm

Center distance between lamps

70mm±2mm

Material

Interior material

stainless steel

Shell material

stainless steel

Humidification mode

Electrical heating and steam humidification

Control system

Operation interface

Digital intelligent, touch input (be able to program control)

Operation mode

Program control / constant operation

Setting range

According to temperature range

Input

BPT (black-panel thermometer), PT-100 sensor

Subsidiary function

Upper and lower limits alarm, alarm display, timer function.

Standard parts

20 stainless steel sample holder

Safety device

Electric leakage, overload power-off protection, over temperature protection, water shortage, ground protection.

Power supply

220V±10%,50Hz

Power (kw)

3

Share to: